Electrical Conductivity
Conductivity Meter 411
- Remove cover. Turn meter on.
- Make sure mode is microS/cm.
- Make sure an A appears in upper left corner by pressing auto read.
- Calibrate.
- Leave sensor in air. Press CAL. (a 1 appears - lower right.)
- Let stabilize. An [/-A] appears in upper left corner.
- Place sensor in 1413 solution. Make sure sensor is submerged.
- Press CAL. (a 2 appears - lower right.)
- Let stabilize. An [/-A] appears.
- Rinse sensor thoroughly.
- Measure sample.
- Place sensor in sample. Make sure sensor is submersed.
- Press READ.
- Let stabilize. An [/-A] appears.
- Take reading.
- If microS/cm => move decimal 3 places to left.
- If mS/cm=> take direct reading.
- Record with 2 decimals.
- Rinse between samples.
- Leave probe in air.
- Turn off meter.
- Cover.